Features
Precise Temperature Control System:
1. The SLAN system adopts Peltier technology which can accurately and quickly control increases and decreases of temperature with a precision of ±0.1℃ and features long service life, no noise and no pollution, etc.;
2. Multiple-point temperature control technology ensures the uniformity between different wells. The temperature correction technology guarantees the accuracy of temperature and thus ensures the accuracy of assay result;
3. PCR tube temperature-control technology guarantees the accuracy of actual temperature of reagent in the tube and enables the SLAN to easily adjust for different reactions using different volumes.
Sensitive Photoelectric Detection System
1. High sensitivity photoelectric detector with no background noise which is fast and stable;
2. Lifetime maintenance-free ultra-bright LED cold light source with a large signal value and high stability;
3. Unique optical fiber conduction technology that greatly improves fluorescence collection efficiency and avoids fluorescence interference between neighboring tubes;
4. The system automatically selects best gain value for each tube without the need for user configuration. The SLAN®has an extremely broad measuring range for fluorescence;
5. Extremely high detection sensitivity with an extremely low fluorescence background value. It’s not necessary for users to make optical correction or background correction.
SLAN-96P Multi-tasking Software Interface
(analysis screen)
Unique Electronic Automatic Hot Lid System
1. Flexible heating technology ensures uniform hot lid temperature;
2. The electronic automatic hot lid has an adjustment function for different 0.2ml reaction tubes to ensure even tube pressure;
3. Once the assay starts, the hot lid has automatic locking function to prevent assay failure following accidental opening.
Other functions
· Power-off protection | · Automatic saving result | · Export of result |
· Real-time data display | · Quantitative and qualitative analysis | · File encryption |
· Automatic baseline optimization | · Digital filtering | · Search function |
· Report sheet print and edit | · Melt curve function |
Specifications
Product model | SLAN-96P |
Sample capacity | 96 wells (dual 48/48 reaction module) |
Sample size | 15-50mL |
Consumable | 0.2mL PCR tubes, 8-tube strips, 48-well plates |
Parallel running mode | Dual reaction blocks/running 2 tests independently |
Temperature range | 4-99°C |
Max ramp rate | 4.0°C/sec |
Temperature accuracy | ±0.1°C |
Temperature uniformity | ±0.1°C |
Thermal cycling system | Peltier-based system |
Temperature control mode | Tube control/block control |
Light Source | LED (maintenance-free) |
Detector | High sensitivity photoelectric sensor |
Sensitivity | 1 copy |
Linearity range | 100-1010 |
Resolution | Can discriminate between 1000 copies and 2000 copies |
Excitation | CH1 470nm CH2 530nm CH3 580nm CH4 630nm CH5/CH6 custom-made |
Emission | CH1 510nm CH2 565nm CH3 620nm CH4 665nm CH5/CH6 custom-made |
Dyes and probes | CH1 FAMTM /SYBRGreen® CH2 VIC®/HEX/J0ETM /TET CH3 ROX/TexasRed® CH4 CY5TM |
Hot-lid | 30°C~108°C (default105°C, adjustable) electronic automatic hot-lid |
Power supply | 230VAC, 50Hz |
Power consumption | 850VA |
Dimensions | 380mm x 520mm x 250mm (WxDxH) |
Weight | 18Kg |
Communication | RS232, USB |
Qualification | CE |